Skip to main content

Table 1 Device comparison. Some characteristics that are not identical in the two devices are given. Most other characteristics are similar, for example NO scrubbing of inhaled air and the exhalation flow control.

From: Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children

 

NIOX

NIOX MINO

Measurement range (ppb)

1.5–200

5–300

Accuracy

< 50 ppb: ± 2.5 ppb

≥ 50 ppb: ± 5%

< 50 ppb: ± 5 ppb

≥ 50 ppb: ± 10%

Visual feedback

Computer screen

Display via mirror

Calibration on site

Yes

No

Dimensions (cm)

50 × 30 × 40

24 × 13 × 10

Weight (kg)

40

0.8