Reference values for exhaled nitric oxide (reveno) study

Abstract

Background

Despite the widespread use of fractional exhaled nitric oxide (FENO) as a biomarker of airways inflammation, there are no published papers describing normal FENO values in a large group of healthy adults.

Objective

The aim of this study was to establish adult FENO reference values according to the international guidelines.

Methods

FENO was measured in 204 healthy, non-smoking adults with normal spirometry values using the on-line single-breath technique, and the results were analysed chemiluminescently.

Results

The main result of the study was the significant difference in FENO values between men and women, thus indicating that gender-based reference FENO values are necessary. The FENO levels obtained at expiratory flows of 50 ml/s ranged from 2.6 to 28.8 ppb in men, and from 1.6 to 21.5 ppb in women.

Conclusion

We propose reference FENO values for healthy adult men and women that could be used for clinical and research purposes.

Background

The presence of nitric oxide (NO) in exhaled air was first described in 1991 by Gustafsson et al.[1], and this was soon followed by a number of publications reporting high fractional concentrations of orally exhaled NO (FENO) in subjects with various pulmonary diseases [2]. FENO is generally measured on line by having the subject blow directly into the analyser and obtaining immediate results [3], but breath can also be collected remotely into inert bags and analysed subsequently (off line) [3].

Although the pathophysiological meaning is still unclear [4], it has been demonstrated that NO levels in exhaled air are higher in asthmatics than in healthy subjects, increase during spontaneous or induced asthma exacerbations, and decrease after anti-inflammatory treatment [5].

Many studies have clearly demonstrated that a number of factors can affect FENO values, and so the European Respiratory Society (ERS) and American Thoracic Society (ATS) established particular recommendations for exhaled and nasal NO measurements in 2005 [6] in order to allow the comparison of data from different research centres.

Clinicians and researchers seeking to apply FENO measurements in everyday practice are obviously interested in knowing what are normal FENO values in healthy subjects, but very few attempts have been made to establish such reference values, and experimental findings are usually only compared with those observed in the healthy controls recruited for any particular study. Buchvald et al. [7] have recently found that upper normal FENO levels in children aged 4–17 years ranged from 15 parts per billion (ppb) in the youngest to 25 ppb in adolescents, with a mean increase of 1 ppb per year. To the best of our knowledge, there are no published studies indicating similar reference values for adults.

The aim of this study was to establish reference adult FENO values according to the international guidelines.

Materials and methods

Study subjects and protocol

This open-label study was conducted in three Italian centres (Brescia, Parma and Verona) and recruited local medical school students and colleagues, who were given a short description of the project, and the inclusion and exclusion criteria.

Healthy subjects were defined as individuals with normal spirometry values and without a history of any significant diseases. Furthermore, in accordance with the ATS/ERS guidelines [6], particular care was taken to avoid the known confounding factors that may affect FENO measurements: in particular, smokers and ex-smokers were excluded; none of the volunteers was taking any drug or medication or had experienced a recent upper or lower airways infection, and none reported any clinical manifestation of allergic diseases or positive skin prick tests for common inhalant allergens.

The study was approved by the Ethics Committee of each centre and all of the participants gave their written informed consent.

Fractional exhaled NO measurements

Fractional exhaled NO (FENO) was measured using a chemiluminescence analyser (CLD88, Ecomedics, Switzerland) whose lower and upper limit of detection (LOD) was respectively 0.06 ppb and 100 ppb. The same type of instrument was used at all of the centres, and was calibrated at 0 and 100 ppb as recommended by the manufacturer.

FENO was measured in accordance with international guidelines [6]. Briefly, after inhaling to total lung capacity, the subjects exhaled through a mouthpiece equipped with a 0.2-μm pore size bacterial filter into an exhalation circuit consisting of an ultrasonic flow meter, one-way valve and one sampling port. NO was sampled directly in the analyser (at a flow rate of 250 ml/min) through a Teflon side arm tube attached to the sampling port. The sampling tube was 60 cm long with an internal diameter of 1/8 of in. Both expiratory flow and FENO values were simultaneously displayed on a computer attached to the analyser. FENO was measured before the subjects underwent spirometry.

Different expiratory flow rates were ensured by placing expiratory resistors (Breath kit, Sievers Instruments, USA) in the exhalation circuit, which yielded expiratory flow rates of 50, 100 and 200 ml/s. The subjects were asked to exhale at a constant flow, which they could readily see displayed on the computer screen in the form of a bar that remained red until target flow was obtained, and then turned green; if the flow dropped below or increased above the desired range, the green bar changed back to red. Although the target expiratory flows were strictly controlled and maintained during the expiration, a tolerance of ± 10% was considered acceptable, and the exhalation continued until a stable plateau had been reached.

Three FENO plateau measurements varying by <10% were made at each flow rate, and the average value was recorded. As the subjects inhaled ambient air, its NO concentration was measured at the time of each test and, if high (>30 ppb), the data were discarded. The influence of ambient NO levels was further excluded by placing an NO-scrubbing filter in the inspiratory limb of the collection apparatus. The data were stored on a computer and analysed using NO analysis software.

Spirometry

The patients underwent spirometry using a spirometer connected to a computer for data analysis (Vmax 22, Sensor Medics, Yorba Linda, CA, USA), and FEV1 and FVC were measured in accordance with the ATS standard procedure [8].

Data analyses

We first analysed the three subgroups of subjects from each centre and then all of the subjects as a whole. As there was no significant difference between the two analyses, for the sake of simplicity, we shall here describe only the results of the first.

Spearman's correlation test was used to verify the correlations between the variables. Between-group comparisons were made using non-parametric analysis of variance (Kruskal-Wallis test) and, if significant, the Mann-Whitney U test (M-W test). Logarithmic transformation was applied to the NO values in order to normalise the curve and the groups were compared using ANOVA; however, in order to simplify the reading, the data are presented as their original values and analysed non-parametrically. Bonferroni's correction for multiple tests was applied.

In the multivariate analysis of the odds ratio estimates, logistic regression was carried out backwise with pre-assigned P values of > 0.05 controlling step removal; the model was evaluated using three goodness-of-fit chi-square statistics.

All of the analyses were made using SPSS Rel. 13.0 statistical package (SPSS Inc., Chicago, IL).

Results

Table 1 shows the demographic data, physical and spirometric parameters, and FENO values. The demographic data, physical parameters, and mean spirometric and FENO values of the healthy non-smoking subjects studied in the three centres were pooled as there were no significant between-centre differences (data not shown).

Table 1 Demographic data, physical parameters and FENO values in studied subjects. Mean values and standard deviation (SD).

Table 2 shows the distribution of the FENO values. Of the 204 recruited subjects (male/female ratio = 1), 78 (38%) were aged 19–30, 65 (32%) were aged 31–40, 39 (19%) were aged 41–50, and 22 (11%) were aged 51–60 years. There was no significant difference in age between the sexes.

Table 2 Data distribution of fractional exhaled nitric oxide values.

All of the subjects underwent spirometry and FENO measurement at an exhaled flow of 50 mL/s, and respectively 178 (92 men and 86 women) and 179 subjects (93 men and 86 women) also had FENO measured at the exhaled flows of 100 and 200 mL/s. Twenty-five of the 26 subjects who were unable to perform the FENO procedure at the exhaled flow of 200 mL/s were also unable to do so at 100 mL/s.

There were significant gender-related differences in body mass index (BMI), height, weight and body surface area (BSA), forced expiratory volume at the first second (FEV1), and forced vital capacity (FVC), but no gender-related difference in exhaled flow values.

At all of the studied flow, FENO levels were significantly lower in the women than in the men (Table 1). FENO levels did not correlate with age, lung function or anthropometric values.

There was a positive correlation between the FENO values at the different exhaled flows: FENO50 vs. FENO100 r = 0.82, p < 0.001; FENO50 vs. FENO200 r = 0.74, p < 0.001; FENO100 vs. FENO200 r = 0.9, p < 0.001.

FENO levels were not correlated with age (r = 0.1, p = 0.21, Spearman's test) or with lung function or anthropometric values. FENO levels at all studied flows were significantly lower in females than those observed in men (Table 1).

Logistic regression analysis was performed considering sex as dependent variable and centre, weight, height, age, FEV1, FVC, FENO50, FENO100 and FENO200 as potentially predictive factors.

Weight and FVC were identified as predictive variables able to distinguish between males and females (data not reported).

Discussion

The primary aim of this study was to measure FENO in a population of healthy controls aged 19–65 years at a flow rate of 50 mL/s using the on-line single breath technique.

Figure 1
figure 1_485

Fractional exhaled nitric oxide (FENO) levels in men and women at three expiratory flows. Mean values and 95% confidence intervals.

We also analysed FENO at flow rates of 100 and 200 mL/s in order to obtain their normal FENO values, and compare the ability of normal adults to expire at such different flows.

The main finding was that, even after adjusting for age, height, weight, BMI and BSA, FENO values were significantly lower in women at all of the studied expiratory flows, which means that different gender-related reference values need to be applied.

FENO levels at the most frequently used expiratory flow rate of 50 ml/s was 2.6–28.8 ppb in men and 1.6–21.5 ppb in women. The FENO levels at 50 ml/s usually reported in studies of healthy adults fall within the 10–20 ppb range [9] but, as in the case of other biological parameters, we observed some individuals with unexplained higher or lower levels despite our strict study inclusion and exclusion criteria. It may therefore be more prudent to define normal FENO values in terms of percentiles, and we would suggest considering the fifth and 95th percentiles (4.5–20.6 ppb for males, and 3.6–18.2 ppb for females), as references for healthy subjects, and taking further diagnostic and clinical steps in the case of subjects whose FENO levels fall outside this range.

Gender-related differences in adult FENO levels were first reported by Jilma et al.[10], who examined the concentrations of exhaled NO and plasma nitrate, and were confirmed by Tsang et al.[11] in a cohort of 121 healthy non-smoking subjects, and by van der Lee et al.[12] However, our data were collected in accordance with the most recent guidelines. It is not clear why this difference exists, but Grasemann et al.[13] have shown that it is partly associated with the NO synthase 1 genotype in healthy females; factors related to hormone production are less plausible, as Morris et al.[14] have shown that there is no temporal relationship between the measurements of NO production and urinary sex steroid conjugates during the menstrual cycle, thus suggesting that estrogens do not modulate FENO concentrations.

We speculate that a further possible reason is the difference in airway surface area and calibre [15, 16]. The same flow rate in airways of different calibres may differently dilute NO, which moves by means of gaseous diffusion into a smaller lumen (i.e. in females), thus leading to a lower NO concentration. Brooks et al.[17] have demonstrated that there is no within-gender correlation between tracheal size and body size or maximal expiratory flows, thus suggesting that the differences in the airway sizes of men and women are true gender-related difference and not simply due to differences in lung or body size. This hypothesis is in line with the findings of Buchvald et al. [7] showing no difference in FENO levels between boys and girl of the same age, but a significant and positive relationship between FENO and age (which leads to a progressive increase in airway surface) in both sexes; if the hypothesis is confirmed, it could be concluded that the low FENO levels in women may simply be an artefact due to the use of a constant exhaled flow rate rather than a real reduction in NO airway production. In this regard, Nguyen et al. [18] have recently shown that the measurement of both FENO and nitrogen oxides (NOX) in exhaled breath condensate is more indicative of airway NO production than FENO alone; further studies should be carried out to verify whether there are any sexual differences in exhaled NOx.

Our data confirm that FENO values are inversely related to the exhalation flow rate [19], and demonstrated that an exhaled flow of 50 mL/s was feasible in all our subjects, which is in line with the published guidelines [6].

We also measured FENO levels at higher expiratory flows of 100 and 200 ml/s, because it has been suggested that extended exhaled NO measurements can distinguish alveolar and bronchial inflammation [20, 21]. However, we found that there was a strong positive correlation (r = >0.7) between FENO levels at different expiratory flows, and that it was not always possible to obtain reproducible expirations at higher flows in our healthy subjects. Further studies of large numbers of patients with proximal and distal airway inflammation should therefore be carried out in order to evaluate whether the FENO measurements at different expiratory flow rates may lead to information that is as useful as that obtained at 50 ml/s.

The mean FENO levels at different expiratory flows in our subjects are comparable with those previously reported by some authors in healthy non-smoking subjects[18, 21], but slightly lower than those reported by others, particularly those observed at the expiratory flow rate of 50 ml/s [23]. In this regard, Borrill et al.[24] have recently compared the FENO levels measured using three different commercially available analysers and found significant differences between them. This raises the important question of variability between NO analysers. This is an important point as Muller et al. [25] have recently shown that the main factors responsible for the different NO readings provided by different analysers are differences in calibration gases and procedures. Our study was not intended to compare the NO readings provided by different analysers, but we are confident of the reliability of our results because Borrill et al.[24] found that the most reproducible data was that obtained using the CLD88, probably because it has the lowest detection limit and fastest response time, and because it is CE MDD approved and totally compliant to the standard required by the ATS/ERS recommendations.

In conclusion, our study demonstrated that measuring FENO measurement at an expiratory flow rate of 50 ml/s was feasible in a population of 204 healthy subjects aged 19–65 years, and indicates that different normal FENO values should be defined for males and females.

Abbreviations

ATS:

American Thoracic Society

CE MDD:

European Community Medical Device Directive

FENO :

fractional exhaled NO

ERS:

European Respiratory Society

NOX :

nitrogen oxides

Ppb:

parts per billion

References

  1. Gustafsson LE, Leone AM, Persson MG, Wiklund NP, Moncada S: Endogenous nitric oxide is present in the exhaled air of rabbits, guinea pigs and humans. Biochem Biophys Res Commun 1991, 181:852–7.

    Article  CAS  PubMed  Google Scholar 

  2. Kharitonov SA: Exhaled markers of inflammatory lung diseases: ready for routine monitoring? Swiss Med Wkly 2004, 134:175–92.

    CAS  PubMed  Google Scholar 

  3. Kissoon N, Duckworth LJ, Blake KV, Murphy SP, Taylor CL, Silkoff PE: FE(NO): relationship to exhalation rates and online versus bag collection in healthy adolescents. Am J Respir Crit Care Med 2000, 162:539–45.

    Article  CAS  PubMed  Google Scholar 

  4. Ricciardolo FL, Sterk PJ, Gaston B, Folkerts G: Nitric oxide in health and disease of the respiratory system. Physiol Rev 2004, 84:731–65.

    Article  CAS  PubMed  Google Scholar 

  5. Kharitonov SA, Barnes PJ: Biomarkers of some pulmonary diseases in exhaled breath. Biomarkers 2002, 7:1–32.

    Article  CAS  PubMed  Google Scholar 

  6. ATS/ERS Recommendations for standardized procedures for online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide. 2005Am J Respir Crit Care Med 2005, 171:912–30.

  7. Buchvald F, Baraldi E, Carraro S, Gaston B, De Jongste J, Pijnenburg MW, et al.: Measurements of exhaled nitric oxide in healthy subjects age 4 to 17 years. J Allergy Clin Immunol 2005, 115:1130–6.

    Article  CAS  PubMed  Google Scholar 

  8. Clausen JL, Coates AL, Quanjer PH: Measurement of lung volumes in humans: review and recommendations from an ATS/ERS workshop. Eur Respir J 1997, 10:1205–6.

    Article  CAS  PubMed  Google Scholar 

  9. Silkoff PE, Carlson M, Bourke T, Katial R, Ögren E, Szefler SJ: The Aerocrine exhaled nitric oxide monitoring system NIOX is cleared by the US Food and Drug Administration for monitoring therapy in asthma. J Allergy Clin Immunol 2004, 114:1241–56.

    Article  CAS  PubMed  Google Scholar 

  10. Jilma B, Kastner J, Mensik C, Vondrovec B, Hildebrandt J, Krejcy K, Wagner O, et al.: Sex differences in concentrations of exhaled nitric oxide and plasma nitrate. Life Sci 1996, 58:469–76.

    Article  CAS  PubMed  Google Scholar 

  11. Tsang KW, Ip SK, Leung R, Yung E, Tipoe Gl, Chan SL, Shum IH, et al.: Exhaled nitric oxide: the effects of age, gender and body size. Lung 2001, 179:83–91.

    Article  CAS  PubMed  Google Scholar 

  12. van der Lee I, van den Bosch JM, Zanen P: Reduction of variability of exhaled nitric oxide in healthy volunteers. Respir Med 2002, 96:1014–20.

    Article  CAS  PubMed  Google Scholar 

  13. Grasemann H, van's Gravesande KS, Bûscher R, Drazen JM, Ratjen F: Effects of sex and of gene variants in constitutive nitric oxide synthases on exhaled nitric oxide. Am J Respir Crit Care Med 2003, 167:1113–6.

    Article  PubMed  Google Scholar 

  14. Morris NH, Sooranna SR, Steer PJ, Warren JB: The effect of the menstrual cycle on exhaled nitric oxide and urinary nitrate concentration. Eur J Clin Invest 1996, 26:481–4.

    Article  CAS  PubMed  Google Scholar 

  15. Franklin PJ, Taplin R, Stick SM: A community study of exhaled nitric oxide in healthy children. Am J Respir Crit Care Med 1999, 159:311–4.

    Article  Google Scholar 

  16. Latzin P, Beck J, Griese M: Exhaled nitric oxide in healthy children: variability and a lack of correlation with atopy. Pediatr Allergy Immunol 2002, 13:37–46.

    Article  PubMed  Google Scholar 

  17. Brooks LJ, Byard PJ, Helms RC, Fouke JM, Strohl KP: Relationship between lung volume and tracheal area as assessed by acoustic reflection. J Appl Physiol 1988, 64:1050–4.

    Article  CAS  PubMed  Google Scholar 

  18. Nguyen TA, Woo-Park J, Hess M, Goins M, Urban P, Vaughan J, et al.: Assaying all of the nitrogen oxides in breath modifies the interpretation of exhaled nitric oxide. Vascul Pharmacol 2005, 43:379–84.

    Article  CAS  PubMed  Google Scholar 

  19. Silkoff PE, McClean PA, Slutsky AS, Furlott HG, Hoffstein E, Wakita S, et al.: Marked flow-dependence of exhaled nitric oxide using a new technique to exclude nasal nitric oxide. Am J Respir Crit Care Med 1997, 155:260–7.

    Article  CAS  PubMed  Google Scholar 

  20. Mahut B, Delacourt C, Zerah-Lancner F, De Blic J, Harf A, Delclaux C: Increase in alveolar nitric oxide in the presence of symptoms in childhood asthma. Chest 2004, 125:1012–8.

    Article  CAS  PubMed  Google Scholar 

  21. Condorelli P, Shin HW, George SC: Characterizing airway and alveolar nitric oxide exchange during tidal breathing using a three-compartment model. J Appl Physiol 2004, 96:1832–42.

    Article  CAS  PubMed  Google Scholar 

  22. Brindicci C, Ito K, Resta O, Pride NB, Barnes PJ, Kharitonov SA: Exhaled nitric oxide from lung periphery is increased in COPD. Eur Respir J 2005, 26:52–9.

    Article  CAS  PubMed  Google Scholar 

  23. Vints AM, Oostveen E, Eeckhaut G, Smolders M, De Backer WA: Time-dependent effect of nitrate-rich meals on exhaled nitric oxide in healthy subjects. Chest 2005, 128:2465–70.

    Article  CAS  PubMed  Google Scholar 

  24. Borrill Z, Clough D, Truman N, Morris J, Langley S, Singh D: A comparison of exhaled nitric oxide measurements performed using three different analysers. Respir Med 2006, in press.

    Google Scholar 

  25. Muller KC, Jorres RA, Magnussen H, Holz O: Comparison of exhaled nitric oxide analysers. Respir Med 2005, 99:631–7.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

We would like to thank Olga Acampa, Lucia Dalle Vedove and Beatrice Ragnoli for their valuable contribution in subject recruitment.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Mario Malerba.

Additional information

Competing interests

The author(s) declare that they have no competing interest.

Authors' contributions

OM: substantial contribution to study conception and design, sample collection, data acquisition, analysis and interpretation; involved in drafting the article.

GT: substantial contribution to study conception and design, data analysis and interpretation, critically reviewing the draft for important intellectual content.

MC: substantial contribution to study conception and design, sample collection; involved in drafting the articles.

LP: substantial contribution to study conception and design, data interpretation and statistical analysis; involved in drafting the article; final approval of the version to be published.

AM: substantial contribution to study conception and design, data interpretation and statistical analysis; involved in drafting the article; final approval of the version to be published.

CT: substantial contribution to study conception and design, data interpretation and statistical analysis; involved in drafting the article; final approval of the version to be published.

MM: substantial contribution to study conception and design, sample collection, critically reviewing the draft for important intellectual content.

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Olivieri, M., Talamini, G., Corradi, M. et al. Reference values for exhaled nitric oxide (reveno) study. Respir Res 7, 94 (2006). https://0-doi-org.brum.beds.ac.uk/10.1186/1465-9921-7-94

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/1465-9921-7-94

Keywords